National University Corporation - Notice of request for submission of materials1 set of Nuclear Magnetic Resonance (NMR) System

This procurement is covered by the WTO Agreement on Government Procurement, Japan-EU Economic Partnership Agreement or Japan-UK Comprehensive Economic Partnership Agreement.

Japanese

Publishing date Sep 11, 2023
Type of notice Notice of request for submission of materials
Procurement entity National University Corporation - Shimane
Classification
0024 Professional/Scientific & Controlling Instruments & Apparatus
Summay of notice ⑴ Classification of the products to be procured : 24
⑵ Nature and quantity of the products to be purchased : 1 set of Nuclear Magnetic Resonance (NMR) System
⑶ Type of the procurement : Purchace
⑷ Basic requirements of the procurement :
A The superconducting magnet (SCM) of the NMR system must generate a magnetic field of 11.74 Tesla (corresponding to 500MHz) and be capable of being transported and carried with liquid cryogen (liquid helium and nitrogen) in it.
B The minimum retention period for liquid helium and nitrogen in the SCM must be 150 days and 15 days, respectively. The SCM must be equipped with a cryogen reclamation system to control evaporated cryogen for both helium and nitrogen.
C The frequency band of the NMR spectrometer must be 460-535 MHz for the high frequency (HF), and 10-210 MHz for low frequency (LF).
D The temperature variable range in the NMR spectrometer must be between -170 ℃ to 250 ℃, and compressed air must be available for NMR measurements above 80 ℃.
E The spectrometer must be available for both solid and liquid samples.
F The NMR detector for liquids must have a diameter of 5 mm and be capable of detecting nuclei of 1H, 19F, 15N-31P, 39K, and 109Ag.
G The NMR detector for solids must have a diameter of 3.2 mm and be capable of detecting nuclei of 1H, 19F, 31P, 7Li, 11B, 23Na, 27Al, 13C, 79Br, 207Pb, 29Si, 6Li, and 15N.
H The system must be equipped with an automatic sample changer with at least 24 slots, which must be available for both solid and liquid NMR measurements.
I The system must have a second workstation for remote control and operation of the system.
⑸ Time limit for the submission of the requested material : 17 : 00 13 October, 2023