National University Corporation - Notice of request for submission of materialsHigh-Resolution Mass Spectrometer 1 Set

This procurement is covered by the WTO Agreement on Government Procurement, Japan-EU Economic Partnership Agreement or Japan-UK Comprehensive Economic Partnership Agreement.

Japanese

Publishing date Dec 26, 2024
Type of notice Notice of request for submission of materials
Procurement entity National University Corporation - Shizuoka
Classification
0024 Professional/Scientific & Controlling Instruments & Apparatus
Summay of notice ⑴ Classification of the products to be procured : 24
⑵ Nature and quantity of the products to be purchased : High-Resolution Mass Spectrometer 1 Set
⑶ Type of the procurement : Purchase
⑷ Basic requirements of the procurement :
A The analyzer must adopt a quadrupole time-of-flight (QTOF) system equipped with ion mobility separation mechanisms.
B The ionization system must include Electrospray Ionization (ESI), and to allow measurement of unstable samples, the spray needle should not be charged with voltage.
C For improved maintenance, the ESI system must employ an orthogonal configuration.
D The system must support Matrix-Assisted Laser Desorption/Ionization (MALDI) and allow analysis using both ESI and MALDI with software-based switching.
E To acquire high-speed, high-quality, and highly sensitive data, the laser used for MALDI must be a solid-state laser adjustable in the range of 1 to 10,000 Hz.
F A post-ionization mechanism utilizing a two-stage laser irradiation process must be provided.
G The ion mobility separation mechanism must calculate Collision Cross-Section (CCS) values.
H The system must enable imaging mass spectrometry analysis through ion mobility separation, allowing the separation of isomers from each other and target peaks from noise peaks.
I The mass accuracy of the system must be within 0.8 ppm (with internal standards) and 2 ppm (with external standards).
J The mass resolving power must be 60,000 or greater (measured by full width at half maximum).
K To reduce sensitivity loss during analysis, the capillary in the ion introduction section must be made of glass.
L To minimize mass errors caused by temperature fluctuations, the system must have a mechanism that uses a chiller for control.
M The system must not require liquid helium for maintenance or operation.
⑸ Time limit for the submission of the requested material : 17 : 00 27 January, 2025
⑹ Contact point for the notice : WADA Tetsuaki, Accounting Division, Hamamatsu University School of Medicine, 1-20-1 Handayama Chuo-ku Hamamatsu city 431-3192 Japan, TEL 053-435-2553