National University Corporation - Notice of request for submission of materialsBiochemistry and immunological testing system 1 Set

This procurement is covered by the WTO Agreement on Government Procurement, Japan-EU Economic Partnership Agreement or Japan-UK Comprehensive Economic Partnership Agreement.

Japanese

Publishing date Dec 11, 2023
Type of notice Notice of request for submission of materials
Procurement entity National University Corporation - Miyazaki
Classification
0022 Medical/Dental/Surgical & Veterinary Equipment
0031 Medical Instruments & Apparatus
Summay of notice ⑴ Classification of the products to be procured : 22, 31
⑵ Nature and quantity of the products to be purchased : Biochemistry and immunological testing system 1 Set
⑶ Type of the procurement : Purchase
⑷ Basic requirements of the procurement :
A Biochemistry analyzer-one unit
⒜ Must be capable of performing simultaneous multi-item analysis using a discrete method.
⒝ The instrument should process a maximum of 5,400 tests per hour for the photometric method and maximum of 1,800 tests per hour for the electrolytic method.
B Specimen Pretreatment Conveyance System-one unit
⒜ Must be able to connect to multiple biochemical autoanalyzers and immunochemiluminescence analyzers.
⒝ The system should be capable of processing more than 500 specimens per hour at a 1 : 1 dispensing ratio.
⒞ The dispensing area should be able to detect liquid levels.
⒟ It should be able to dispense 200-800 〓 at once.
⒠ The dispensing area of the system should have droplet prevention function.
⒡ The system can store more than 400 processed specimens.
⒢ It should be able to replenish consumables even during operation.
⒣ It should be selectable of the right/left direction where the specimens are conveyed to.
⒤ It should have a backup device.
C Immunoassay system-one unit
⒜ The system should have a processing capability of 200 tests per hour.
⒝ Sample capacity should be more than 150 samples.
⒞ It should have an automatic re-examination function.
⒟ IT should have a backup device with equal performance to the main unit.
D Seismic isolator device-one unit
⒜ The device should be installed with seismic isolation.
⒝ The height of the seismic isolator should be less than 10 mm.
⒞ It should withstand a superimposed load of more than 500 kg per square meter.
⒟ It does not need regular maintenance.
E Temperature Control System-one unit
⒜ It can use cloud servers.
⒝ The operational status of the equipment can be checked in real time.
⒞ Data should be downloadable in CSV format.
F Benchtop Centrifuge-one unit
⒜ Maximum rotation per minute should be more than 3,500 rpm. The maximum centrifugal force should be more than 2,300×g.
G Floor-Standing Refrigerated Centrifuge-one set
⒜ Maximum rotation per minute should be more than 4,800 rpm. The maximum centrifugal force should be more than 4,150×g.
⑸ Time limit for the submission of the requested material : 17 : 00 26 January, 2024
⑹ Contact point for the notice : Takanori Kurogi, Procurement Section, Accounting Division, National University Corporation University of Miyazaki, 1-1 Gakuen Kibana-dai Nishi Miyazaki-shi 889-2192 Japan, TEL 0985-58-7574