National University Corporation - Notice of request for submission of materialsCryogenic Transmission Electron Microscope 1 Set

This procurement is covered by the WTO Agreement on Government Procurement, Japan-EU Economic Partnership Agreement or Japan-UK Comprehensive Economic Partnership Agreement.

Japanese

Publishing date Jul 06, 2026
Type of notice Notice of request for submission of materials
Procurement entity National University Corporation - Miyagi
Classification
0024 Professional/Scientific & Controlling Instruments & Apparatus
Summay of notice ⑴ Classification of the products to be procured : 24
⑵ Nature and quantity of the products to be purchased : Cryogenic Transmission Electron Microscope 1 Set
⑶ Type of the procurement : Purchase
⑷ Basic requirements of the procurement :
① A cryogenic transmission electron microscope.
A It should be equipped with a cold field emission gun, which produces a highly coherent electron beam. The energy spread is supposed to be less than 0.4 eV without using a monochromator.
B The accelerating voltage should be higher than 300 kV.
C The specimen can be tilted more than ±70°and the automated specimen exchange system should be equipped. It must be able to contain up to 12 grids at the same time.
D The energy of electrons contributing to images and diffraction patterns should be selected by using an in-column type energy filter.
E The microscope should have the phase plate system to enhance the small-angle scattered electron image contrast.
F The system should contain a highly sensitive direct electron detector for image recording. It has 24 megapixels (5,760 × 4,092) field of view and the readout speed of 1,500 full frames per second. It should also have a tough camera for electron beam irradiation suitable for recording diffraction patterns with more than 4k × 4k pixels.
② A vacuum evaporator
This is needed to prepare amorphous carbon layer for grid preparation.
A The evacuation should be done by a turbomolecular pump and reachable vacuum degree should be higher than 1 × 10 Pa.
③ A glow discharger
This is needed to hydrophilize the surface of the grid for grid preparation.
A The size of the electrostatic shield electrode should be larger than 50 mm in its diameter.
④ An automated grid plunging device
This is needed to perform plunge-freezing for vitrification of specimens.
A The temperature of the specimen chamber should be controllable at the range from 4 ℃ to 60 ℃.
B The container should keep 1ℓ liquid nitrogen for one hour.
⑸ Time limit for the submission of the requested material : 17 : 00 6 August, 2026
⑹ Contact point for the notice : Hiroaki Kawamura, Procurement Services Office, Finance Department, Tohoku University, 2-1-1 Katahira Aoba-ku Sendai-shi 980-8577 Japan TEL 022-217-4869